| SAPPHIRE UV-VIS Variable Wavelenght
Detector
is a detector with continuously
variable wavelength in the range of 190 - 800 nm (190 - 600 nm) and a
noise level ± 0.5 x 10-5 AU. The unit may be used for routine
analytical chromatography as well as for preparative chromatography,
according to the selected cell. The unit is easy to operate, reference
and sample signals are available for detector diagnostics as well as
information on lamp operating hours. Output signal is available in
both digital and analogue form on the output connector. It is possible
to control the unit using the keypad or an RS232 interface. Connecting
capillaries are led from the detector's right side panel. The detector
is equipped with a high standard deuterium lamp in a special socket
that enables easy changing. Wavelength can be programmatically changed
during the performance of an analysis. The detector performs automatic
wavelength calibration after the lamp has been switched on. A HPLC 04
analytical cell is included as a component of the detector. |
 |
| OPAL UV-VIS Fixed Wavelenght
Detector
is a detector with an interference
filter. By means of this filter and deuterium lamp spectrum can be
this unit adjusted at one fixed wavelength in range of 190 - 800 nm.
Noise level at 254 nm is ± 0.5 x 10-5 AU with an empty cell. The unit
may be used for routine analytical chromatography as well as for
preparative chromatography, according to the selected cell. The unit
is easy to operate, reference and sample signals are available for
detector diagnostics as well as information on lamp operating hours.
Output signal is available in both digital and analogue form on the
output connector. It is possible to control the unit using the keypad
or an RS232 interface. |
 |
| TOPAZ Dual UV Detector
is a new type of UV detector, which
allows measuring absorbance of two wavelengths simultaneously in one
cell in range 200 - 320 nm (increments 1 nm). This unit is used in
liquid chromatography to verify analyzed samples by means of a second
wavelength or in situations when some peaks absorb on different
wavelengths. It is possible to use the detector in analytical,
preparative or flash applications by using different cells. Noise
level at 254 nm is ± 1 x 10-4 AU with an empty cell. The unit is easy
to operate, reference and sample signals are available for detector
diagnostics as well as information on lamp operating hours. Output
signal is available in both digital and analogue form on the output
connector. It is possible to control the unit using the keypad,
external logic signals or an RS232 interface. |
 |
| LCD 2073A Dual UV Detector
This unit has the same features as
the TOPAZ Dual UV Detector but it does
not have a keypad and display thus it is possible to control it by
using
external logic signals or an RS232 interface only. |
 |
| FLASH 06 SINGLE and DUAL Detector
is an OEM build-in detector with one
interference filter (FLASH 06 Single Detector) or two interference
filters (FLASH 06 Dual Detector). The FLASH 06 Dual Detector makes to
possible measuring of two wavelenghs simultaneously. Detector can be
adjusted using interference filter at one or two fixed wavelength in
range of 200 - 600 nm. Noise level at 254 nm is ± 2 x 10-5 AU with an
empty cell. High standard deuterium lamp in a special socket enables
easy changing. Lamp work hours are counted using the built-in counter.
This unit is used in liquid chromatography in flash and preparative
applications. By using analytical cell can be detector used also for
analytical purpose. The cell is easy to replace from the side of the
detector. The unit is designed as build-in unit with 24 V power supply
and RS 232 controlling. On request can be sold in classical case with
230/115 V powering. |
 |
| FLASH 06 DAD Detector FLASH 06 DAD
Detector
is an OEM build-in type of UV diode
array detector (DAD), which allows
measuring absorbance of four wavelengths simultaneously in one cell
just
as measuring of whole spectrum (scan). Scan of whole spectrum is
possible in each time.
This unit is used in liquid chromatography to verify analyzed samples
by
means of four wavelengths or in situations when some peaks absorb on
different wavelengths. It is possible to use the detector in flash and
preparative applications. By using analytical cell can be detector
used also
for analytical purpose. The cell is easy to replace from the side of
the
detector.
Noise level at 254 nm is ± 5 x 10-5 AU with an empty cell.
Wavelength setting is from 200 up to 400 nm in increments of 1 nm.
Lamp work hours are counted using the built-in counter.
The unit is designed as build-in unit with 24 V power supply and RS
232
controlling. On request can be sold in classical case with 230/115 V
powering. |
 |
| Shodex RI-101 Refractive Index
Detector analytical
Differential Refractive Index
Detectors are used for detecting little or no UV active samples such
as sugars or polymers. The Shodex RI-101, employs a number of
improvements for shortening the time for stabilization. By using a
micro-processor it is capable of automatic start-up from eluent
replacement to checking the base-line stabilization. The detector is
fitted with external input and output terminals as well as RS232C
communication ports, permitting advanced automation by linking the
detector with various HPLC systems, enabling advanced automation by
combining the detector with various chromatographs. Main features:
- superior base line stability
- intuitive operating using
- easily overview of operational
conditions and baseline trend
- automatic start-up function for
reference cell filling and baseline stabilization checking
- HELP function for easy validation
and calibration
- control mode (by keypad, or RS232)
- optimized temperature control
method guarantee quick base line stabilization
- leak sensor stop the pump by
solvent leakage
- various automated function
|
 |